- 고선량율 원격강내조사의 코발트-60 이동선원에 의한 선량특성
- Dose Characteristics by the Co-60 Source Oscillations in High Dose Rate After Loading Irradiations
- ㆍ 저자명
- 최태진,김옥배,노홍균
- ㆍ 간행물명
- 의학물리
- ㆍ 권/호정보
- 1990년|1권 1호|pp.51-60 (10 pages)
- ㆍ 발행정보
- 한국의학물리학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Dose distributions around Co- 60 moving source in high dose rate remote afterloading unit, Buchler 3K unit, were experimented with X-omat V films and calculations. In our experiments, film dosimetries have achieved to evaluated the axial dose distributions for source oscillations were 0, 3.5, 5.0 and 6.0 cm in periodically, In results, the dose distributions in axial of source movement showed apparently higher than in transverse direction caused by source locations, dwelling time and air gap in the applicator. In the calculations, the dose rate was derived by using the inverse square law, filteration corrections and Meisberger constant for scatter corrections as source movings. In our experiments and calculations, the average dose uncertainties were showed -2.1$pm$1.9% in fixed sourdce, -2.9$pm$1.8%, -7.4$pm$6.1% and -6.7$pm$4.6% at 3.5 cm, 5.0 cm and 6.0 cm source oscillations, but the calculations have showed very close to experimental dose rate within 4 cm distance from source.