- 공기와 지방에 의한 자기공명영상 인공물에 관한 실험연구
- ㆍ 저자명
- 박경주
- ㆍ 간행물명
- 대한방사선의학회지
- ㆍ 권/호정보
- 1992년|28권 3호|pp.321-326 (6 pages)
- ㆍ 발행정보
- 대한영상의학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
To investigate the nature of artifacts caused by air and chemical shift by fat, a test tube phantom study was performed in 2.0T and 0.5T superconducting magnetic resonance imaging systems in various imaging sequences. Spatial mismapping and signal change made image distortion in spin echo images of the air-column and severs signal voids in gradient echo images. Chemical shift artifacts appeared in the planes of frequency-encoding axis as crescent-shaped signal voids and summated bright signals. Both artkfacts were more prominent in high field strength. Images at 0.5T showed more severe artifacts in the T2-weighted images than those in the other sequences. As the acquisition matrix increased, truncation artifacts decreased and susceptibility artifacts in gradient echo images diminished. Familiarity with causes, appearances and mechanisms of various artifacts will help radiologists for interpretation of MRI.