- 6 MV 광자선조사면내 투과성필터에 의한 조직선량
- Scatter Dose in soft tissue using the partial attenuation filter for 6 MV X-ray of linear accelerator
- ㆍ 저자명
- 최태진,김옥배
- ㆍ 간행물명
- 의학물리
- ㆍ 권/호정보
- 1993년|4권 1호|pp.55-71 (17 pages)
- ㆍ 발행정보
- 한국의학물리학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Measured and calculated the TMR and SMR factors from percent depth dose underpartial attenuators which cover the whole part of the radiation beam with variousfilter thickness from 0 to 50 mm. This study was performed for x-ray beams generated with a 6 MV linear acceleratorat source to surface distance of 100cm in a water phantom for Lipowitz metal. TMR(0,d,t) was derived from non-linear polynomial regression with field sizedifferencies and a given filter thickness. In this experiments, the TMR(0,10,50) of 50mm of filter thickness was showed13.6 % higher than that of open field and SMR(5,10,50) was 38.5% smaller than thatof open field in same depth.