- 조영증강 급속 스핀에코 T1 강조영상: 조영증강 병소의 대조도 향상
- ㆍ 저자명
- 최성욱
- ㆍ 간행물명
- 대한방사선의학회지
- ㆍ 권/호정보
- 1997년|37권 3호|pp.379-384 (6 pages)
- ㆍ 발행정보
- 대한영상의학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Purpose:The purpose of this study was to evaluate the effect of contrast improvement of enhancing brains lesions by inherent magnetization transfer effect in turbo spin-echo(TSE) T1-weighted MR imaging. Materials and Methods:Twenty-six enhancing lesions of 19 patients were included in this study. Using a 1.0T superconductive MR unit, contrast-enhanced SE T1-weighted images(TR=600 msec, TE=12 msec, NEX=2, Acuisition time=4 min 27 sec) and contrast-enhanced TSE T1-weighted images(TR=600 msec, TE=12 msec, NEX=2, Acuisition time=1 min 44 sec) were obtained. Signal intensites at enhancing lesions and adjacent white matter were measured in the same regions of both images. Signal-to-noise ratio(SNR) of enhancing lesions and adjacent white matter, and contrast-to-noise ratio(CNR) and lesion-to-background contrast (LBC) of enhancing lesions were calculated and statistically analysed using the paired t-test. Result:On contrast-enhanced TSE T1-weighted images, SNR of enhancing lesions and adjacent white matter decreased by 18%(p<0.01) and 32%(p<0.01), respectively, compared to contrast-enhanced SE T1-weighted images. CNR and LBC of enhancing lesions increased by 16%(p<0.05) and 66%(p<0.01), respectively. Conclusion:Due to the proposed inherent magnetization transfer effects in TSE imaging, contrast-enhanced T1-weighted TSE images demonstrated a statistically significant improvement in CNR and LBC, compared to conventional contrast-enhanced T1-weighted SE images, and scan time was much shorter.