- 핵심국제비교용 물의 삼중점 셀 제작과 비교 실험 및 불확도 평가
- ㆍ 저자명
- 강기훈,김용규,감기술,Kang. Kee-Hoon,Kim. Yong-Gyoo,Gam. Kee-Sool
- ㆍ 간행물명
- 센서학회지
- ㆍ 권/호정보
- 2004년|13권 2호|pp.114-120 (7 pages)
- ㆍ 발행정보
- 한국센서학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
The $Comet{acute{e}}$ Consultatif de $Thermom{acute{e}}trie$ (CCT) under the Comete International des Poids et Measures (CIPM) has decided to perform the Key Comparison (KC) for triple point of water cells used as a reference fixed point of thermometry at the 21st meeting held at November 2001, and the Bureau International des Poids et Measures (BIPM) has been nominated as a KC coordinator. According to the KC protocol prepared by BIPM, KRISS performed the KC experiments and evaluate a uncertainty. The temperature difference between two reference cells for the Korea Research Institute of Standards and Science (KRISS) and a test cell for the transfer standard, which is moved to BIPM was 0.024 mK and the combined standard uncertainty evaluated 0.055 mK.