- CCT-ICP-MS의 대기분진내 미량원소분석에 대한 적용성
- ㆍ 저자명
- 임종명,이진홍,서만철
- ㆍ 간행물명
- 한국대기환경학회지
- ㆍ 권/호정보
- 2004년|20권 3호|pp.397-409 (13 pages)
- ㆍ 발행정보
- 한국대기환경학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
There has been few studies of either domestic or international to apply CCT-ICP-MS for the precise analysis of As and Cr components associated with airborne particulate matter. To date, the use of CCT-ICP-MS is strongly recommended for the accurate analysis of the toxic trace metals; this is because CCT-ICP-MS technique prevents polyatomic spectral interferences involved in the determination of As and/or Cr components. Taking advantage of CCT-ICP-MS technique, the measurements of about 20 metals were undertaken in this study. The standard reference material (NIST SRM 2783) was used for analytical quality control. To improve analytical accuracy and of acid efficiency, we selected nitric acid based on a test of three kinds of acid for microwave digestion method 1 ) nitric acid. 2) nitric acid and hydrogen peroxide. and 3) nitric acid and perchloric acid. When this method was employed, relative errors to SRM values of Al, As, Cr Fe, Mg, Mn, Pb, Sb, V, and Zn fell below 20%, while those or Ca, Si, and Ti were higher than 20%. The overall results of our study show that the concentrations of As and V determined by CCT-ICP-MS were satisfied with the certificated values within a relative error of 20e1c, whereas those determined by ICP-MS were 10 times higher than the certificated values.