A sensor and measuring system were implemented to measure both uroflow and urophonography signal during urination for diagnose the lower urinary tract symptom. The implemented system was composed of the uroflow sensor using the load-cell, the urophonography measurement sensor so as stethoscope type, pre-processing part for sensors signal detection, amplifier and filter, system control parts and PC measurement program. A simulator of the lower urinary system that is experimental equipment implemented for evaluate the developed system. The two signals were measured using implemented system and analyzed these signal by means of time domain and frequency domain for extraction of the characteristic parameter which can most effectively reflected by the occlusion of the lower urinary system. Furthermore two signals were measured and analyzed with the subject of 5 healthy adult for clinical application possibility of the implemented system. As a result, the most influence bandwidth of effect according to occlusion of the lower urinary system is $253{sim}282$ Hz of the urophonography signal.