- 완충용액유속, 효소.기질 반응 및 전극봉 요인에 따른 발효공정 제어용 흐름주입식 효소센서의 전류값 특성
- ㆍ 저자명
- 송대빈,정효석,김성태,Song. Dae-Bin,Jung. Hyo-Seok,Kim. Sung-Tae
- ㆍ 간행물명
- 바이오시스템공학
- ㆍ 권/호정보
- 2007년|32권 6호|pp.455-461 (7 pages)
- ㆍ 발행정보
- 한국농업기계학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
The electric current of a flow injection type enzyme-sensor was measured to confirm the stable operating conditions of the sensor. The current of the sensor was decreased as the buffer solution velocity increased. Under the limitation of the cycle time to be below 10 minutes, the effective ranges of the buffer solution velocity were suggested $0.10{sim}0.26$, $0.12{sim}0.24$, $0.1{sim}0.25$ and $0.05{sim}0.10;cm/s$ of 1.0, 1.4, 2.4 and 3.4 mm of the electrode diameters, respectively. As the reaction time of the enzyme and the substrate was increased, the current was decreased because of the dilution between the sample and buffer solution. Therefore, it could be recommended that the reaction time was able to be selected as shortly as possible in consideration of the total cycle time. As the result of the experiments using a different volume ratio of the enzyme to substrate, it was concluded that the substrate had to be mixed with the same amount of the enzyme. The current have increased remarkably in proportion to the electrode diameter under 0.1 cm/s of the buffer solution velocity but there was no difference over 0.1 cm/s of the buffer solution velocity. The cross type arrangement of the electrode was highly suggested for application and machining of the sensor.