- 유방촬영(乳房撮影)의 방사선량(放射線量)에 관(關)한 연구(硏究)
- ㆍ 저자명
- 최종학,전만진,김영일,최종운,Choi. Jong-Hak,Jeon. Man-Jin,Kim. Young-Ill,Choi. Jong-Woon
- ㆍ 간행물명
- 대한방사선기술학회지
- ㆍ 권/호정보
- 1981년|4권 1호|pp.31-36 (6 pages)
- ㆍ 발행정보
- 대한방사선과학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
We studied radiation dose in mammography through 34-46 kv range using acryl phantom. The obtained results were as follows: 1. Incident radiation was maximum with high kvp and thin added filtration. 2. Transmitted radiation by acryl phantom and its thickness were in reciprocal relationship. 3. The acryl thickness to produce comparable film density with soft tissue of breast was 6 cm. 4. The X-ray exposure for comparable density radiographs increased mammographic film more than medical x-ray film and the amount of x-ray exposure was directly proportional to the added filtration of x-ray beam. 5. The surface dose of x-ray exposure needed to produce film density of 1.0 for 6cm acryl phantom was 1,084-1,575mR in mammographic film and 476-625 mR in medical x-ray film.