- 뇌 전산화단층촬영시의 수정체 피복선량에 관한 연구
- ㆍ 저자명
- 이은혜
- ㆍ 간행물명
- 대한방사선의학회지
- ㆍ 권/호정보
- 1984년|20권 1호|pp.33-36 (4 pages)
- ㆍ 발행정보
- 대한영상의학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
Radiation exposure dose for patients during brain CT scan with General Electric C.T./T.8800 was measured using CaF2:Mn thermoluminescant dosimeter chips. The skin dose in single slice scan was 1.26$pm$0.10 rad, and that in 9 slice scan was 2.06$pm$0.16 rad So the ratio was 1.63. The surface dose in profile along axial distance in multiple slice scan was 0.27$pm$0.05 rad, too small to induce any pathologic condition. As noted from the above data, the radiation exposure dose during CT scanning is not high compared with other radiologic examinations.