- 진단용(診斷用) X선(線)의 반가층(半價層)에 관(關)한 연구(硏究)
- A Study on Measurement of the Half Value Layer in Diagnostic X-ray
- ㆍ 저자명
- 고신관,Ko. Shin-Gwan
- ㆍ 간행물명
- 대한방사선기술학회지
- ㆍ 권/호정보
- 1984년|7권 1호|pp.53-66 (14 pages)
- ㆍ 발행정보
- 대한방사선과학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
The quality of continuous x-ray beam depends upon the half value layer which varies according to the geometric conditions, the filtration thickness, and the amount of accelerated voltage (KVP). Experiments were conducted on the amount of electric energy that was changed to x-ray energy, and on the relationship between KVp and the intensity of x-rays. The results were as follows: 1. The amount of x-rays were not equal under the condition of the same exposure factor. 2. The intensity of x-rays was attenuated by an exponential function the geometric conditions were "good" and it was not when they were "poor". 3. The thicker the total filtration substance was and the higher the KVp was, the bigger the amount of x-ray energy was. 4. The homogeneity of medium energy x-ray was the best, when the total filtration substance was 3.9mm A1. 5. The mean energy of continuous x-ray was about 45% of KVp.