The general features and measured results of the 143cm, variable-energy, three sector-focused MC-50 cyclotron installed at Korea Cancer Center Hospital are described. The MC-50 cyclotron is designed to produce beams of protons, deuterons, He-3 and alpha particles of maximum energies 50MeV, 25MeV, 66MeV and 50MeV respectively to be used for neutron therapy and radioisotope production. The azimuthal field variation is produced by three sets of spiral ridges having a maximum spiral angle of $55^{circ}$. The RF system, a two-dee quarter-wave system is designed to provide a continuously variable frequency from 15.5 to 26.8 MHz. The first external beam was obtained in January 1986. Subsequent internal ana external beam studies with protons and alphas show a well-behaved beam through the whole beam transport system.