- Digital 분진계와 개인용 분진 포집기의 측정값의 관계
- ㆍ 저자명
- 윤영노,이정주,정호근,최호춘,Yoon. Young-No,Lee. Jeong-Joo,Chung. Ho-Keun,Choi. Ho-Chun
- ㆍ 간행물명
- 예방의학회지
- ㆍ 권/호정보
- 1987년|20권 2호|pp.255-260 (6 pages)
- ㆍ 발행정보
- 대한예방의학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
The digital dust indicator (Sibata P-5), one of the direct·reading instruments was evaluated for the respirable dust in the underground coal mine environments. As a reference, respirable dust was determined using three cyclones and/or impactors. All the tests were performed on aerosol in twenty underground coal mines. The coefficients of mass-relative concentration were $0.067{pm}0.054$ (Mean$pm$Standard deviation) (range: 0.006-0.172). The relationship between relative concentration and temperature was not significant statistically. Also, the relationship of relative concentration and relative humidity was not significant. Mass concentration and relative concentration were $5.31{pm}5.22mg/m^3$ and $162{pm}163$ CPM ($Mean{pm}Standard$ deviation) respectively. The range of mass concentration was $1.22-22.69mg/m^3$; relative concentration 16-628 CPM. The relationship of mass concentration and relative concentration was not significant in these ranges.