- 치과 구내 X-선 촬영기의 사용연한에 따른 해상능에 관한 연구
- A STUDY OF THE RESOLUTION OF DENTAL INTRAORAL X-RAY MACHINES
- ㆍ 저자명
- 김선주,정현대,Kim. Seon Ju,Chung. Hyon De
- ㆍ 간행물명
- 齒科放射線
- ㆍ 권/호정보
- 1990년|20권 2호|pp.289-297 (9 pages)
- ㆍ 발행정보
- 대한구강악안면방사선학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
The purpose of this study was to assess the resolution and focal spot size of dental X-ray machines. Fifty dental X-ray machines were selected for measuring resolution and focal spot size. These machines were used in general dental clinics. The time of installation of the X-ray machine varies from 1 year to 10 years. The resolution of these machines was measured with the test pattern. The focal spot size of these machines was measured with the star test pattern. The following results were obtained: 1. The resolution of dental intraoral X-ray machines was not significantly changed in ten years. 2. The focal spot size of dental intraoral X-ray machines was not significantly increased in ten years. The statistical analysis between the mean focal spot size and nominal focal spot size was significant at the 0.05 level about the more than 3 years used machines.