- 흉부(胸部) X선검사시(線檢査時) 환자(患者)의 피폭선양(被曝線量)에 대(對)한 연구(硏究)
- A Study of the Patient Dose in Chest Radiography
- ㆍ 저자명
- 김창균,Kim. Chang-Kyun
- ㆍ 간행물명
- 대한방사선기술학회지
- ㆍ 권/호정보
- 1990년|13권 1호|pp.3-9 (7 pages)
- ㆍ 발행정보
- 대한방사선과학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
A study was carried out to investigate the technical factors and the patient dose (entrance and absorbed dose) in chest P-A radiography based on the 86 hospitals in Seoul from July 1 to July 30, 1989. As a result of this study, main finding were as follow : 1. 51.2% of the surveyed hospitals made use of $60{sim}69;kVp$ as tube voltage in chest radiography 2. The majority of the surveyed(88.3%) have the use of $6{sim}20;mAs$ as tube current-time. 3. Percentage absorbed doses in patient were showed more than 90 percent in every tube voltage. 4. Object densities were all much the same in all tube voltages. 5. 48.8% of surveyed entrance doses ranged from $100;{mu}Sv$ to $190;{mu}Sv$, and the mean dose was $158;{mu}Sv$.