- ML-6M선형 가속기에서의 BEAM 특성에 대한 고찰
- ㆍ 저자명
- 문언철,윤병운,오양환,Mun. Eon-Cheol,Yun. Byeong-Un,O. Yang-Hwan
- ㆍ 간행물명
- 大韓放射線治療技術學會誌
- ㆍ 권/호정보
- 1992년|5권 1호|pp.115-119 (5 pages)
- ㆍ 발행정보
- 대한방사선치료학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
The beam characteristics and dosimetric measurements of the 6MV X-ray and 6MeV electron beam from a ML-6M linear accelerator are examined. The Percent Depth Dose(PDD) table and the tissue Maximum Ratio(TMR) table are taken from measurement as a function of the field size and the depth. The calculated TMR table from PDD table is compared with those from measurement. Other beam characteristics such as output factor, beam profile(including flatness, symmetry and penumbra), wedge, and the variation of Dmax are presented. All of these dosimetric measurements sufficiently characterized the beam to permit safe clinical use.