  • PERFORMANCE TEST FOR A PDS MICRODENSITOMETER MODEL 1010GMS
  • PERFORMANCE TEST FOR A PDS MICRODENSITOMETER MODEL 1010GMS
저자명
Hong. S.S.,Paek. W.G.,Lee. S.G.
간행물명
天文學會誌
권/호정보
1992년|25권 1호|pp.23-46 (24 pages)
발행정보
한국천문학회
파일정보
정기간행물|ENG|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

기타언어초록

The electrical, mechanical and optical capabilities have been tested of the microdensitometer PDS 1010GMS at the Korea Astronomy Observatory. The highest stage of scan speed 255 csu (conventional speed unit) is measured to be 47 mm/s. At this speed the position is displaced by $4{mu}m$ to the direction of scanning and the density is underestimated by $0.4{sim}0.7D$. Standard deviation in the measured density is proportional to $A^{-0.46}$, where A is the area of scan aperture. The accuracy of position repeatability is ${pm}1{mu}m$, and that of density repeatability is ${pm}(0.003{sim}0.03)D$. Callier coefficient is determined to be 1.37; the semispecular density is directly proportional to the diffuse density up to 3.5D. Because the logarithmic amplifier has a finite response time, the densities measured at high scan speeds are underestimated to the degree that speeds higher than 200 csu are inadequate for making an accurate astronomical photometry. After power is on, an about 5 hour period of warming is required to stabilize the system electrically and mechanically as well. On the basis of this performance test, we have determined the followings as the optimum scan parameters for the astronomical photometry: For the scan aperture $10;sim;20{mu}m$ is optimal, and for the scan speed. $20;{sim};50$ csu is appropriate. These parameter values are chosen in such a way that they may keep the density repeatability within ${pm}0.01D$, the position displacement under $1{mu}m$, and the density underestimation below 0.1D even in high density regions.