- 건습시약 스트립을 사용한 임상화학분석장치 개발[I]
- ㆍ 저자명
- 유동주,정태화,민홍기,허웅,Yu. Dong-Ju,Jeong. Tae-Hwa,Min. Hong-Gi,Heo. Ung
- ㆍ 간행물명
- 의공학회지
- ㆍ 권/호정보
- 1992년|13권 4호|pp.299-306 (8 pages)
- ㆍ 발행정보
- 대한의용생체공학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
For quantitative measurement of reflected light from a clinical diagnostic scrip, a prototype of reflectance photometer was designed. The strip loader and cassette were made to obtain more accurate re(leclance parameters. The strop was illuminated at $45^{circ}C$ through optical fiber and the intensity of reflected light was determined at rectangulat angle using a photodiode. The kubelka-munk coefficient and reflection optical density were determined ar four different wavelengths(500,550,570 and 610nm) for blood glucose strip. For higher concentration than 300mg/dl about glucose, a saturation state of absorbance was observed at 500,550 and 570nm. The correlation between glucose concentration and parameters was the best at 610nm.