  • 이종조직 판막의 내구성
저자명
김종환
간행물명
大韓胸部外科學會誌
권/호정보
1992년|25권 5호|pp.494-503 (10 pages)
발행정보
대한흉부외과학회
파일정보
정기간행물|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

기타언어초록

The durability of the xenograft cardiac substitute valves is of a great concern on the clinical grounds. Four groups of tc tal and consecutive patients to the end of study operated on between 1976 and 1984 were Group ISM, 291 patients of MVR, ISA, 65 patients of AVR, and ISMA, 107 patients of MVR+AVR with the standard Ionescu-Shiley bovine pericardial valve, and H, 163 patients of valve replacement with the Hancock porcine aortic vlave. Operative mortality was 5.2%[ISM], 10.8%[ISA], 7.5%[ISMA] and 6.1%[H]. Early survivors were followed up for a total of 1148.3 patient-years[pt-yrs] [ISM], 271.2 pt-yrs [ISA], 488.1 pt-yrs[ISMA] and 822.9 pt-yrs[H]. Linearized late mortality was 2.1% /pt-yr [ISM], 1.l%/pt-yr[ISA], 1.8%/pt-yr[ISMA] and 1.8% /pt-yr[H]. Thromboembolic complication was experienced at the linearized rate of 1.045% /pt-yr [ISM], 1.475%/pt-yr[ISA], 0.615%/pt-yr[ISMA] and 1.822%/pt-yr[H], and bleeding complication at the rate of 0.871% /pt-yr[ISM], 0.63% /pt-yr[ISA], 0.205% /pt-yr [ISMA] and 0.729%a /pt-yr[H], respectively. Prosthetic valve endocarditis occurred at the rate of 0.610% /pt-yr[ISM], 1.475% /pt-yr[ISA], 1.639% /pt-yr[ISMA] and 0.972% /pt-yr[H]. The linearized annual incidence of primary tissue failure was 1.655%/pt-yr[ISM], l. 475%/pt-yr[ISA], 1.639% /pt-yr[ISMA], 2.187% /pt-yr[H] and 1.785% /pt-yr[Group HM : MVR with Hancock valve]. The incidence of tissue failure was significantly high in the patients younger than 30 years of age compared with the older patients. The actuarial survival was 87.7$pm$2.5% at 10 years[ISM], 94.3$pm$3.2% at 11 years[ISA], 89.6$pm$3.4% at 10 years[ISMA] and 81.3$pm$6.6% at 12 years[HM], The freedom from thromboembolism was 93.2$pm$2.0% at 10 years[ISM], 90.6$pm$4.6% at 11 years[ISA], 95.8$pm$2.6% at 10 years[ISMA] and 80.9$pm$11.1% at 12 years[HM], And, the freedom from primary tissue failure was 84.2$pm$3.8% and 28.1$pm$23.0% at 9 and 10 years[ISM], 60.4$pm$16.9% at 11 years[ISA], 62.3$pm$12.7 at 10 years[ISMA] and 65.6$pm$9.8% at 12 years[HM]. In conclusion, the standard Ionescu-Shiley and the Hancock bioprosthetic valves are excellent in their antithrombogenicity and long-term survival. However, the features of the structural failure with the prolonged follow-up beyond 10 years appear to be guarding, and the clinical indications of these bioprostheses seem to be quite limited.