- 진단방사선 영역에서 피폭선량 감소를 위한 기술적 연구 - 관전압과 부가여과판을 중심으로 -
- ㆍ 저자명
- 하호영,Ha. Ho-Young
- ㆍ 간행물명
- 대한방사선기술학회지
- ㆍ 권/호정보
- 1992년|15권 1호|pp.89-97 (9 pages)
- ㆍ 발행정보
- 대한방사선과학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
X-ray quality is identified numerically by half value layer(H.V.L) and the HVL is affected by the kVp and the amount of filtration in the useful beam. X-ray quality evaluated by H.L.D is influenced by kVp and filtration. Author had several experiments with phantom in diameter of 8 cm normal adult chest, for reduction of radiation dose of the patients in diagnostic radiology and got some results. 1. H.V.L is increased the thicker the filter and the higher the kVp. 2. If the kVp is increased from 60 to 120, the skin dose can be reduced as 34%(Skin dose of 60 kVp with 4 mmAl filter : 100%). 3. If the 4 mmAl filter with 60 kVp is added to x-ray tube, skin dose can be reduced as 23% than no filter. 4. Therefore high kVp and filtration can increase output to input dose ratio and 120 kVp and 4 mmAl filter were most effective for reduction of patient dose in chest radiography.