- 확산차이를 유도한 실험모델들에서의 확산강조자기공명영상을 이용한 연구
- ㆍ 저자명
- 서대철,문치웅,이대근,서대건,최기영,고기영,이호규,박평환,임태환
- ㆍ 간행물명
- 대한방사선의학회지
- ㆍ 권/호정보
- 1995년|33권 2호|pp.165-170 (6 pages)
- ㆍ 발행정보
- 대한영상의학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Purpose : To see the stability and error in the diffusion-weighted magnetic resonance (MR) imaging technique in the experimental models and to observe the signal in the early cerebral lesions of the animal models.Materials and Methods : Diffusion coefficients of acetone and distilled water were measured by diffusion-weighted MR image and were compared with actual values. Differentiation of diffusion from perfusion were done at the resin flow phantom. The signal intensities caused by early parenchymal changes were measured in normal, hypovolemic, and embolic, and dead animal model by using diffusion-weighted image and compared with pathologic finding and vital staining.Results : Diffusion coefficients of acetone and distilled water were $4.48{ imes}10^{-3}$ which were very close to the actual values. Diffusion-weighted MR image obtained at flow phantom was not affected by flow(perfusion) at the 100-400 of b-factor range. Animal study done at that b-factor range revealed a significant signal difference between the left and right sides only at the embolic model induced by polyvinyl alchol particles (p<0.05). These changes were not detected in microscopic finding but could be identified in vital staining.Conclusion : Diffusion-weighted MR image can be used to detect early parenchymal change when the appropriate b-factor range was applied.