  • 확장된 SSD에 기인한 Electron beam의 Output 및 특성 변화에 관한 연구
  • A Study on clinical Considerations caused by inevitably Extended SSD for Electron beam therapy
저자명
이정우,김정만,Lee. Jeong-U,Kim. Jeong-Man
간행물명
大韓放射線治療技術學會誌
권/호정보
1996년|8권 1호|pp.29-35 (7 pages)
발행정보
대한방사선치료학회
파일정보
정기간행물|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

기타언어초록

We are often faced with the clinical situations that is inevitably extended SSD for electron beam therapy due to anatomical restriction or applicator structure. But there are some difficulties in accurately predicting output and properties. In electron beam treatment , unlike photon beam the decrease in output for extended SSD does not follow inverse-square law accurately because of a loss of side scatter equilibrium, which is particularly significant for small cone size and low energies. The purpose of our study is to analyze the output in changing with the energy, cone size, air gap beyond the standard SSD and to compare inverse-square law factor derived from calculated effective SSD, mominal SSD with measured output factor. In addition, we have analyzed the change of PDD for several cones with different SSDs which range from 100cm to 120cm with 5cm step and with different energies(6MeV, 9MeV, 12MeV, 16MeV, 20MeV). In accordance with our study, an extended SSD produces a significant change in beam output, negligible change in depth dose which range from 100cm to 120cm SSDs. In order to deliver the more accurate dose to the neoplastic tissue, first of all we recommend inverse-square law using the table of effective SSDs with cone sizes and energies respectively or simply to create a table of extended SSD air gap correction factor. The second we need to have an insight into some change of dose distribution including PPD, penumbra caused by extended SSD for electron beam therapy.