- 비루관 풍선 확장술과 스텐트 삽입술시 사용되는 갈고리의 유용성
- ㆍ 저자명
- 김태형
- ㆍ 간행물명
- 대한방사선의학회지
- ㆍ 권/호정보
- 1996년|35권 1호|pp.33-37 (5 pages)
- ㆍ 발행정보
- 대한영상의학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Purpose : To evaluate the usefulness of a hook developed for pulling out the guide wire in nasolacrimal balloon dilatation and stent placement. Materials and Methods : The head was made of stainless steel wire in a shape of a question mark, and the body was made of four stainless steel pipes. The head was attached to the body in three different ways : A) the head was anterior to the body, B) the head was posterior to the body, and C) the head was angled of 30 degrees forward from the direction of the body. 90 epiphora patients took part in a test to compare and evalvate the usefulness of different types of hook. Average number of trials, success rate and average time of the procedure were recorded. Results : The average trial number for different attachment methods A, B, and C was 4.5, 6.4, and 2.5, respectively, and the success rate 83, 70, and 97%, respectively. The average time of the procedure was 31 minutes. Conclusion : The C type hook was the most effective, and the new technique using the hook was more convenient and time-saving than the old technique using a nasal endoscopy.