- 주파수의존형 전자근관장측정기의 정확도에 관한 연구
- ㆍ 저자명
- 박주현,노병덕,이승종,Park. Ju-Huun,Roh. Byoung-Duck,Lee. Seoung-Jong
- ㆍ 간행물명
- 大韓齒科保存學會誌
- ㆍ 권/호정보
- 1996년|21권 1호|pp.150-160 (11 pages)
- ㆍ 발행정보
- 대한치과보존학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
The present study was to evaluate the accuracy of the frequency dependent type apex locator, Root-ZX. The subjects included 505 root canals of 238 teeth treated by the Department of Conservative Dentistry, and 22 human premolars which were schduled to be extracted for the orthodontic reasons. The results were as follows ; 1. The working lengths determined by Root-ZX were compared with radiographic readings. Of the total 505 root canals, 66 % showed coincidence within ${pm}0.5mm$ and the average readings of Root-ZX were $0.13mm{pm}1.05$ longer than those of radiographic readings. 2. The length difference between the file tip determined by Root-ZX and the apical constriction in extracted teeth were measured. Of the total 24 root canals, 70.8 % showed coincidence within ${pm}0.5mm$ and the average readings of Root-ZX were $0.12mm{pm}0.50$beyond the apical constriction. 3. The vitality of the teeth did not show any statistical difference(p>0.05) in the accuracy of the Root-ZX readings. The presence of the periapical lesions, however, significantly lowered the percentage of ${pm}0.5mm$ accuracy in Root-ZX measurements.(p<0.05). In the presence of periapical lesions, the percentage within ${pm}0.5mm$ was significantly lower.