- SPECT 회전중심 측정에 있어서 콜리메타의 영향
- The Influence of Collimators on SPECT COR Measurements
- ㆍ 저자명
- 이만구,Lee. Man-Koo
- ㆍ 간행물명
- 대한방사선기술학회지
- ㆍ 권/호정보
- 1996년|19권 1호|pp.89-94 (6 pages)
- ㆍ 발행정보
- 대한방사선과학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Misalignment between the electronic and mechanical axes of rotation will result in artifact generation and image degradation during single photon emission computed tomography(SPECT) reconstruction. Acceptance and quality control testing procedures have not emphasized the variability in center of rotation(COR) measurements caused by collimators and the need to verify uniformity across the full collimator field of view (FOV). Variation from the mean COR across the FOV was tested in four different collimators using multiple point source acquisitions. The mean COR was different for each collimator and one of the four had a > 0.5 pixel difference from the mean COR on some area of the FOV, This variation makes this collimator unacceptable for SPECT acquisition. Thus, initial acceptance testing of SPECT collimators should verify a uniform COR across the full FOV and collimators with a variability from the mean COR > 0.5 pixels should be rejected.