- Photodiode를 사용한 측정기의 임상응용을 위한 실험
- Experiment for Clinical Application with Photodiode
- ㆍ 저자명
- 김유현,권수일,허준,Kim. You-Hyun,Kwon. Soo-Il,Huh. Joon
- ㆍ 간행물명
- 대한방사선기술학회지
- ㆍ 권/호정보
- 1996년|19권 2호|pp.71-77 (7 pages)
- ㆍ 발행정보
- 대한방사선과학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Studies were conducted to determine the clinical application of photodiode. We compared photodiode with ion-chamber as to change of tube potential, tube current, mAs and measured decreasing rate of penetration dose. When tube potential was changed from 60 kVp to 120 kVp, output of photodiode and ion-chamber were changed from 0.4 to 1.625, and 1.018 to 4.268, respectively. This was a good agreement to theory that $I=Kv^2it$(I is intensity, K is constant, v is tube potential, i is tube current, t is time). Characteristics for change of tube current and mAs were also a good agreement to theory. And comparison in decreasing rate of penetration dose was similar except above 6 cm in depth. Our results indicated that photodiode was a good instrument for relative measurement of radiation exposure, but we can not use the photodiode for absolute radiation dose.