- 타이태늄 합금의 생체적합도에 관한 연구
- ㆍ 저자명
- 한종현,허성주,구영,최용창,정종평,박중근,Han. Chong-Hyun,Heo. Seong-Joo,Ku. Young,Choi. Young-Chang,Chung. Chong-Pyong,Park. Chung-Keun
- ㆍ 간행물명
- 대한치주과학회지
- ㆍ 권/호정보
- 1998년|28권 3호|pp.401-408 (8 pages)
- ㆍ 발행정보
- 대한치주과학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Screw shaped implants of Titanium-13Zirconium-6Niobium(newly developed), Titanium-6Zirconium-6Sn-6Niobium(newly developed) and Titanium-6Aluminum-4Vanadium were machined with square top and inserted in rabbit bone for 3 months. Biomechanical tests(removal torque) showed Titanium-13Zirconium-6Niobium and Titanium-6Zirconium-6Sn-6Niobium to be more stable in the bone bed than those of Titanium-6Aluminum-4Vanadium. Titanium-13Zirconium-6Niobium implants demonstrated a mean removal torque of 31.59Ncm while Titanium-6Aluminum-4Vanadium demonstrated a mean removal torque of 25.27Ncm and Titanium-6Zirconium-6Sn-6Niobium revealed a mean removal torque of 37.44Ncm and were statistically significance in Wilcoxon Signed Rank test(P<0.05). Histomorphometrical comparisons were performed on $10;{mu}m$ thick undecalcified ground sections in the light microscope and Titanium-13Zirconium-6Niobium showed more mean bone-tometal contact ratio than to other twotitanium alloys but had no statistically significant differences were found among the three materials(P>0.01).