- 컬럼 스위칭 고속액체크로마토그라프법을 이용한 혈장 중 플루코나졸의 분석
- ㆍ 저자명
- 지준필,진숙,이미경,김양배,김종국,Jee. Jun-Pil,Jin. Sook,Lee. Mi-Kyung,Kim. Yang-Bae,Kim. Chong-Kook
- ㆍ 간행물명
- 藥劑學會誌
- ㆍ 권/호정보
- 2000년|30권 1호|pp.51-54 (4 pages)
- ㆍ 발행정보
- 한국약제학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
A column-switching high performance liquid chromatographic method has been developed for the determination of a fluconazole in human plasma. Each plasma sample was centrifuged for 10 min at 5000 g. After an aliqout of the supernatant was taken to nylon microcentrifuge filter, these samples were centrifuged for 10 min at 5000 g. An aliqout of the supernatant was injected directly onto the HPLC column. Deionized water was run for 2 min at a flow rate of 1.0 ml/min to retain fluconazole in an extration column, while proteins and endogenous interferences were eluted to the waste. The analyte was then back-flushed onto an analytical column, $C_{18}$ reversed-phase column. The mobile phase for analytical column, 0.01 M sodium acetate (pH 5.0)-methanol (65:35, v/v), was run at a flow rate of 1.0 ml/min. The column effluent was monitored by ultraviolet detection at 261 nm. The retention time for fluconazole was 11.76 min in human plasma. The detection limit for fluconazole in human plasma was $0.2;{mu}g/ml$. No interference from endogenous substances was observed.