- portable 흉부촬영시 공간산란선량에 관한 연구
- A Study on the Scattered Dose in Portable Chest Radiography
- ㆍ 저자명
- 안봉선,이환형,Ahn. Bong-Seon,Lee. Hwan-Hyung
- ㆍ 간행물명
- 대한방사선기술학회지
- ㆍ 권/호정보
- 2000년|23권 2호|pp.63-67 (5 pages)
- ㆍ 발행정보
- 대한방사선과학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
The purpose of this study is to survey the present status of portable radiography and the result of free space scattered dose rate when taking a radiography at the general hospital or the university hospital in Taejon city. The results were as follows; 1. The number of cases using portable radiography for three years increased to averages 16.2%, 7.7% per year from January 1st in 1996 to December 31st in 1998. 2. The average of distance of adjacent patients was 219.1 cm at the ward. 3. For portable chest radiography, the free space scattered dose rate was 10.5 mSv/hr at 50 cm distance, 1.8 mSv/hr at 100 cm distance, and 0.2 mSv/hr at 200 cm distance. Therefore, in case of portable chest radiography at the ward, the average of distance of adjacent patients is 219.1 cm, so it does not have influence on the adjacent patients. But during the portable radiography, a guardian who is close to the patient, doctor, nurse and radiologic technologists has to set up the shield to prevent from the unnecessary radiation or the distance should be as great as possible from the mobile X-ray equipment.