- 자동 및 수동현상에 따른 Insight 필름과 Ektaspeed Plus 필름의 흑화도 비교
- ㆍ 저자명
- 윤숙자,Yoon. Suk-Ja
- ㆍ 간행물명
- 대한구강악안면방사선학회지
- ㆍ 권/호정보
- 2001년|31권 1호|pp.17-22 (6 pages)
- ㆍ 발행정보
- 대한구강악안면방사선학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Purpose: To compare the film density of Insight dental X-ray film (Eastman Kodak Co., Rochester, NY, USA) with that of Ektaspeed Plus film (Eastman Kodak) under manual and automatic processing conditions. Materials and Methods : Insight and Ektaspeed Plus films were exposed at three different exposure conditions with an aluminum step wedge on the films under the three different exposure times. The exposed films were processed by both manual and automatic ways. The Base plus fog density and the optical density made by exposing step wedge were calculated using a digital densitometer (model 07-443, Victoreen Inc, Cleveland, Ohio, USA). The optical densities of the Insight and Ektaspeed film versus thickness of alumimun wedge at the same exposure time were plotted on the graphs. Statistical analyses were applied for comparing the optical densities of the two films. Results: The film density of both Insight films and Ektaspeed Plus films under automatic processing condition was significantly higher over the manual processing. The film density of Insight films was significantly higher than that of Ektaspeed Plus films on both automatic and manual processing conditions. Conclusion: The radiation exposure time can be reduced when using Insight over Ektaspeed Plus film. To take the full advantage of reducing exposure time, Insight film should be processed automatically.