Comparison Study of the Modulation Transfer Function of a Prototype a-Se based Flat Panel Detector with Conventional Speed Class 400 Film/screen System
[STEP1]서지반출 형식 선택
파일형식
@
서지도구
SNS
기타
[STEP2]서지반출 정보 선택
  • 제목
  • URL
돌아가기
확인
취소
  • Comparison Study of the Modulation Transfer Function of a Prototype a-Se based Flat Panel Detector with Conventional Speed Class 400 Film/screen System
  • Comparison Study of the Modulation Transfer Function of a Prototype a-Se based Flat Panel Detector with Conventional Speed Class 400 Film/screen System
저자명
Park. Jang-Yong,Park. Ji-Koon,Kang. Sang-Sik,Moon. Chi-Woong,Lee. Hyung-Won,Nam. Sang-Hee
간행물명
電子工學會論文誌. Journal of the Institute of Electronics Engineers of Korea. SC, 시스템 및 제어
권/호정보
2003년|40권 3호|pp.163-171 (9 pages)
발행정보
대한전자공학회
파일정보
정기간행물|ENG|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

영문초록

디지털 방사선 검출기의 성능을 측정하기 위해 변조전달할수(Modulation Transfer Function)을 측정하여 기존 방사선 영상 시스템인 필름/스크린과 비교하였다. 제작된 디지털 방사선 검출기는 TFT 패널위에 진공 증착법을 통해 비정질 셀레늄이 코팅되었다. 비교 측정물이 감도 400의 필름/스크린은 아날로그 방식의 방사선 검출기로서 현재 임상에서 이용되고 있는 필름이다. Square wave & slit 측정법을 통해 두 시스템의 변조전달함수를 측정하였다. 필름/스핀린의 반치폭(FWHM)은 357㎛(1.4 lp/mm at 50% spatial frequency)으로 측정되었으며, 디지털 방사선 검출 시스템은 200㎛(2.5 lp/mm at 50%)였다. 영상의 표현성능 평가의 결과 디지털 방사선 시스템은 필름 시스템보다 높은 것으로 나타났다.

기타언어초록

To evaluate the performance of the digital radiography(DR) system developed in our group, the modulation transfer function(MTF) was measured and compared with that of an analog X- ray detector, film/screen system. The DR system has an amorphous selenium(a-Se) layer vacuum-evaporated on a TFT flat panel detector. The speed class 400 film/screen (Fuji) system has been being used in the clinical field as analog X-ray detectors. Both the square wave and slit method were used to evaluate their MTF. The square method was applied to both film/screen and the DR system. The slit method, however, was applied to only DR system. The full-width half maximum resolution of film/screen was 357${mu}{ extrm}{m}$(1.4 lp/mm at 50% spatial frequency), and the resolution of DR was limited to 200${mu}{ extrm}{m}$(2.5 lp/mm at 30%). These results indicate the measured resolution limitations approximate to the pixel pitch, 139 ${mu}{ extrm}{m}$ of TFT. The MTF of DR is higher than that of film/screen by the factor of 1.785. It is proved that our a-Se based DR system has potential usefulness in the clinical field.