- 파노라마방사선사진에서 골형태 계측과 구내표준필름에서 구리당량치의 상관관계
- ㆍ 저자명
- 김재덕,Kim. Jae-Duk
- ㆍ 간행물명
- 대한구강악안면방사선학회지
- ㆍ 권/호정보
- 2003년|33권 3호|pp.131-135 (5 pages)
- ㆍ 발행정보
- 대한구강악안면방사선학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Purpose: To determine whether the mandibular radiomorphometric indices in panoramic radiography are correlated with the bone mineral density of Cu-equivalent images in intraoral film. Materials and Methods: The bone mineral density (BMD) of the mandibular premolar area was measured in the Cu-equivalent image of intraoral film. The Panoramic Mandibular Index (PMI) and Mandibular Cortical Width (MCW) were measured in panoramic radiographs of six dry mandibles, and the Pearson correlation between PMI, MCW, and BMD were tested. Results: There were no significant correlations between PMI and BMD (r = 0.280), nor between MCW and BMD (r =0.237). Conclusion: The results show that PMI and MCW were poor diagnostic indicators of mandibular BMD in the six dry mandibles used in this study. The correlationship between the mandibular radiomorphometric indices (PMI and MCW) and mandibular BMD needs to be researched further using large in vivo patient samples.