- 유동센서 보정용 캘리브레이션 제트 시스템 개발
- ㆍ 저자명
- 장조원,변영환,Chang. J.W.,Byun. Y.H.
- ㆍ 간행물명
- 한국항공운항학회지
- ㆍ 권/호정보
- 2003년|11권 1호|pp.41-55 (15 pages)
- ㆍ 발행정보
- 한국항공운항학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
A calibration jet system using separate blower is developed to calibrate a flow sensor effectively. Designed open circuit type mini calibration jet system, which has the dimension of $0.5m(W){ imes}1.17m(H)$ is small compared with conventional calibration jet systems. The exit of nozzle has exchangeable contractions with a cross section area of $38.5cm^2$ , and a cross section area of $113.1cm^2$, respectively. The ranges of wind speed at exit of exchangeable nozzles are $7.5{sim}42;m/s$ and $1.8{sim}16.5;m/s$, respectively. The input power for the high pressure blower is 1.18kW. The turning vanes for corner was rolled flat plate parallel to the flow direction. The flow conditioning screen was located immediately downstream of the wide-angle diffuser. The honeycomb and two flow conditioning screens were located in the stagnation chamber. From the economical point of view and the simplicity of the calibration jet system set up and handling, it can be said that the developed calibration jet system is an effective calibration jet system. This system can also be used to calibrate the flow sensor with high resolution.