- 전기로 및 봉입형 응고점 셀 내의 온도구배가 미치는 표준백금저항온도계 온도측정의 불확도 요소 평가
- ㆍ 저자명
- 강기훈,감기술,김용규,송창호,Kang. Kee-Hoon,Gam. Kee-Sool,Kim. Yong-Gyoo,Song. Chang-Ho
- ㆍ 간행물명
- 센서학회지
- ㆍ 권/호정보
- 2004년|13권 6호|pp.411-416 (6 pages)
- ㆍ 발행정보
- 한국센서학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
In the international temperature scale of 1990 (ITS-90), standard platinum resistance thermometer (SPRT) is a defining standard thermometer used in the temperature range from 13.8033 K to $961^{circ}C$. Uncertainty of SPRT is about several mK and uncertainty of defining fixed points of the ITS-90 which is used for calibrating SPRT is about several tenth of mK. Above $0^{circ}C$. the defining fixed points are gallium melting point and indium, tin, zinc, aluminium and silver freezing points which are all realized using an electric furnace or a liquid bath. To realize freezing point of tin ($231.928^{circ}C$) and zinc ($419.527^{cir}C$), two 3-zone furnaces which have 3 electric heaters were manufactured. Temperature gradient of the constructed furnaces were tested. Uncertainty caused by temperature gradient of furnace and immersion effect of SPRT in the sealed-type freezing point cells were evaluated 0.038 mK for tin freezing point and 0.036 mK for zinc freezing point.