- o2/Co2 분석기를 사용하여 폐의 기능적 잔기용량을 계측하는 질소세척법 개발
- ㆍ 저자명
- 김군진,김경아,이재헌,이태수,차은종,Kim. Goon-Jin,Kim. Kyung-Ah,Lee. Jae-Hun,Lee. Tae-Soo,Cha. Eun-Jong
- ㆍ 간행물명
- 센서학회지
- ㆍ 권/호정보
- 2004년|13권 5호|pp.329-334 (6 pages)
- ㆍ 발행정보
- 한국센서학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Functional residual capacity (FRC) is an important diagnostic parameter measured using $N_{2}$ analyzer. Since $N_{2}$ analyzer is expensive as well as cumbersome for use of noisy vacuum pump, the FRC measurement becomes possible only in large well-equipped hospitals. The present study introduced a new $TN_{2}$ wash-out technique to measure FRC by $O_{2}/CO_{2}$ analysis, which is relatively cheaper and much simpler to apply. Slower $O_{2}$ response was compensated for high frequency to be coincided with $CO_{2}$ response, thereby enabled indirect, but accurate $N_{2}$ concentration measurement. FRC was estimated by continuous integration of expired $N_{2}$ volume obtained with air flow signal. Experiment with 3 L syringe, a standard calibration device recommended by the American Thoracic Society, demonstrated less than 1% error at 0, 1, and 2 L. Correlation coefficient was almost ideal, guaranteeing linear estimation of FRC. The present technique is inexpensive and simple to apply, thus should he of great convenience.