- Kodak Insight 구내필름의 치아우식증 진단능에 대한 연구
- ㆍ 저자명
- 윤영남,이상래,이병도,Yoon. Young Nam,Lee. Sang-Rae,Lee. Byung-Do
- ㆍ 간행물명
- 대한구강악안면방사선학회지
- ㆍ 권/호정보
- 2004년|34권 1호|pp.19-23 (5 pages)
- ㆍ 발행정보
- 대한구강악안면방사선학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Purpose: To compare the diagnostic accuracy of Kodak Insight film with other intra-oral films in the detection of dental caries. Materials and Methods : Periapical radiographs of 99 extracted human teeth with sound proximal surfaces and interproximal artificial cavities were made on Kodak Ultra speed, Ektaspeed, Agfa Ektaspeed and Kodak Insight films and automatically processed. Six dentists examined the presence of dental caries using a five-point confidence rating scale and compared the diagnostic accuracy by ROC (Receiver Operating Characteristic) analysis and ANOVA test. Results: The sensitivity of Kodak Ultra speed, Ektaspeed, Agfa Ektaspeed and Insight film were 0.84, 0.77, 0.75 and 0.79 respectively. The specificity of Kodak Ultra speed, Ektaspeed, Agfa Ektaspeed and Insight film were 0.97. 0.95. 0.96 and 0.94 respectively. The mean ROC areas (Az) of Kodak Ultra speed, Ektaspeed, Agfa Ektaspeed and Insight film were 0.917, 0.910, 0.894, 0.909 respectively. There was no significant differences between Az of Insight film and other films (p = 0.178). Conclusion: Theses results suggested that Kodak Insight film have the comparative diagnostic accuracy of dental caries with Ultraspeed and Ektaspeed films.