- GC의 주입방식 차에 따른 고농도 악취황 성분의 검량오차 연구 : 주입부피의 고정방식 대비 주입농도의 고정방식 간 비교연구
- ㆍ 저자명
- 김기현,최여진,Kim. Ki-Hyun,Choi. YJ
- ㆍ 간행물명
- 한국대기환경학회지
- ㆍ 권/호정보
- 2005년|21권 2호|pp.269-274 (6 pages)
- ㆍ 발행정보
- 한국대기환경학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
In this work, analytical bias arising from the gas chromatographic determination of sulfur compounds was evaluated by the application of direct loop injection method to the GC/PFPD detection of four sulfur compounds including H$_{2}$S, CH$_{3}$SH, DMS, and DMDS. For the proper evaluation of analytical uncertainties involved in GC calibration, we employed two comparative techniques of calibration at fxed concentration injection (CFCI) vs calibration at fixed volume injection (CFVI) method. The results of our study indicate that CFCI method exhibits very poor sensitivity due to the matrix effect with varying injection volumes. On the other hand, as CFVI method overcomes such limitation, it can be used to obtain very accurate quantification of S compounds at their high concentration levels above a few to a few tens ppb.