- 인공고관절 치환술에서 대퇴주대 회전에 따른 시멘트막 두께 변화
- ㆍ 저자명
- 박용국,김진곤,Park. Yong-Kuk,Kim. Jin-Gon
- ㆍ 간행물명
- 한국정밀공학회지
- ㆍ 권/호정보
- 2007년|24권 2호|pp.140-148 (9 pages)
- ㆍ 발행정보
- 한국정밀공학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
THRA(Total Hip Replacement Arthroplasty) has been widely used for several decades as a viable treatment of otherwise-unsolved hip problems. In THRA surgery, cement mantle thickness is critical to long-term implant survival of femoral stem fixed with cement. Numerous studies reported thin or incomplete cement mantle causes osteolysis, loosening, and the failure of implant. To analyze the effect of femoral stem rotation on cement thickness, in this study, we select two most popular stems used in THRA. Using CAD models obtained from a 3D scanner, we measure the cement mantle thickness developed by the rotation of a femoral stem in the virtual space created by broaching. The study shows that as the femoral stem deviates from the target coordinates, the minimum thickness of cement decreases. Therefore, we recommend development of a new methodology for accurate insertion of a femoral stem along the broached space. Also, modification of the stem design robust to the unintentional movement of a femoral stem in the broached space, can alleviate the problem.