- 비교 측정을 통한 PM-10 질량농도의 오차 분석
- ㆍ 저자명
- 정창훈,조용성,황승만,정용국,유재천,신동석,Jung. Chang-Hoon,Cho. Yang-Sung,Hwang. S.M.,Jung. Y.G.,Ryu. J.C.,Shin. D.S.
- ㆍ 간행물명
- 한국대기환경학회지
- ㆍ 권/호정보
- 2007년|23권 6호|pp.689-698 (10 pages)
- ㆍ 발행정보
- 한국대기환경학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Particular attention is paid to the validity of PM-10 measurement using beta attenuation method. In this study, the inter-comparison studies were undertaken for PM-10 sampler with different measuring methods. The three beta attenuation monitoring instruments (BAM1020, FH C-14, and DIGS-150) and filter based gravitational method using High Volume Air Sampler were collocated and mass concentrations were compared from August to November, 2006. The results of the investigation show that beta attenuation monitoring instruments are comparable with gravitational method within 10% of measurement error. For specific conditions under high humidity and high concentrations, the inter-comparison results also showed a relatively good agreement and this can be explained as the effect of inlet heating of beta attenuation instrument. Subsequently, this study shows that beta attenuation monitoring instruments can be equivalently used with reasonable validity when compared gravimetric method.