Estimating blood pressure using the pulse transit time of the two measuring from pressure pulse and PPG
[STEP1]서지반출 형식 선택
파일형식
@
서지도구
SNS
기타
[STEP2]서지반출 정보 선택
  • 제목
  • URL
돌아가기
확인
취소
  • Estimating blood pressure using the pulse transit time of the two measuring from pressure pulse and PPG
  • Estimating blood pressure using the pulse transit time of the two measuring from pressure pulse and PPG
저자명
김기련,예수영,김재형,전계록,Kim. Gi-Ryon,Ye. Soo-Young,Kim. Jae-Hyung,Jeon. Gye-Rok
간행물명
센서학회지
권/호정보
2008년|17권 2호|pp.87-94 (8 pages)
발행정보
한국센서학회
파일정보
정기간행물|ENG|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

기타언어초록

Blood pressure (BP), one of the most important vital signs, is used to identify an emergency state and reflects the blood flow characteristics of the cardiovascular system. The conventional noninvasive method of measuring BP is inconvenient because patients must wear a cuff on their arm and the measurement process takes time. This paper proposes an algorithm for estimating the BP using the pulse transit time (PTT) of the photoplethysmography (PPG) and pressure pulse from finger at the same time as a more convenient way to measure the BP. After recording the electrocardiogram (ECG), measuring the pressure pulse, and performing PPG, we calculated the PTT from the acquired signals. Then, we used a multiple regression analysis to measure the systolic and diastolic BP indirectly. Comparing the BP measured indirectly using the proposed algorithm and the real BP measured with a sphygmomanometer, the systolic pressure had a mean error of ${pm}3.240$ mmHg and a standard deviation of 2.530 mmHg, while the diastolic pressure had a satisfactory result, i.e., a mean error of ${pm}1.807$ mmHg and a standard deviation of 1.396 mmHg. These results are more superior than existing method estimating blood pressure using the one PTT and satisfy the ANSI/AAMI regulations for certifying a sphygmomanometer i.e., the measurement error should be within a mean error of ${pm}5$ mmHg and a standard deviation of 8 mmHg. These results suggest the possibility of applying our method to a portable, long-term BP monitoring system.