- 표준백금저항온도계를 이용한 온도측정능력 상호비교
- ㆍ 저자명
- 감기술,강주식,이용재,이광복,김용규,박승남,Gam. K.S.,Kang. C.S.,Lee. Y.J.,Lee. K.B.,Kim. Y.G.,Park. S.N.
- ㆍ 간행물명
- 센서학회지
- ㆍ 권/호정보
- 2009년|18권 1호|pp.86-94 (9 pages)
- ㆍ 발행정보
- 한국센서학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Temperature measurement capability was inter-compared using the transfer standard platinum resistance thermometers(SPRT) among four laboratories of KRISS. The transfer SPRTs were primarily calibrated at the triple point of water and Ga melting point, then used at inter-comparison experiment. Temperature difference of calibration value between temperature laboratory and length laboratory at $20^{circ}C$ was -0.7 mK and +2.4 mK at density laboratory. Temperature measured near $20^{circ}C$, $25^{circ}C$ and $30^{circ}C$ at fluid flow laboratory was deviated by $34.2{sim}80.4;mK$ from the calibration values of the transfer SPRT. Ga melting points was inter-compared among three laboratories, and the difference of Ga melting points against the standard Ga melting point of temperature laboratory were $0.03{sim}0.54;mK$ at length laboratory and 0.02 mK at density laboratory.