- 공공설치용 자동제세동기의 사용성 연구
- Usability Study on Public Access Defibrillator
- ㆍ 저자명
- 오영진,Oh. Young-Jin
- ㆍ 간행물명
- 산업경영시스템학회지
- ㆍ 권/호정보
- 2009년|32권 1호|pp.61-66 (6 pages)
- ㆍ 발행정보
- 한국산업경영시스템학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Usability study of the CU Medical "i-PAD" AED was conducted to validate the usability of the device, primarily focused on the trained user as well as the utility of the training material. Comparing these results of the trained participants to those obtained by Callejas et al. (2004), who studied video-trained users of two commercially available Philips AEDs (FR2 and HS1) using a male manikin only, the participants in this study were more successful (100% vs. 85.7%-FR2 vs. 88.9%-HS1) and faster to shock (58.5s-i-PAD vs. 91s-FR2 vs. 90s-HS1). While the focus of this study was on validating ease of use for trained bystanders, the results of the untrained condition showed that, from a usability perspective, these devices perform as well as any other AEDs already approved and available for purchase for untrained bystander use, as evidenced by the comparison to various published research results.