  • Histomorphometry and stability analysis of early loaded implants with two different surface conditions in beagle dogs
  • Histomorphometry and stability analysis of early loaded implants with two different surface conditions in beagle dogs
저자명
Kim. Dong-Seok,Kim. Dae-Gon,Park. Chan-Jin,Cho. Lee-Ra
간행물명
The journal of advanced prosthodontics
권/호정보
2009년|1권 1호|pp.10-18 (9 pages)
발행정보
대한치과보철학회
파일정보
정기간행물|ENG|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

기타언어초록

STATEMENT OF PROBLEM. Despite an improved bone reactions of Mg-incorporated implants in the animals, little yet has been carried out by the experimental investigations in functional loading conditions. PURPOSE. This study investigated the clinical and histologic parameters of osseointegrated Mg-incorporated implants in early loading conditions. MATERIAL AND METHODS. A total of 36 solid screw implants(diameter 3.75 mm, length 10 mm) were placed in the mandibles of 6 beagle dogs. Test groups included 18 Mg-incorporated implants. Turned titanium implants served as control. Gold crowns were inserted 4 weeks after implant placement and the dogs were immediately put on a food diet. Implants were observed for 10 weeks after loading. Radiographic assessments and stability tests were performed at the time of fixture installation, $2^{nd}$ stage surgery, 4 weeks after loading, and 10 weeks after loading. Histological observations and morphometrical measurements were also performed. RESULTS. Of 36 implants, 33 displayed no discernible mobility, corresponding to successful clinical function. There was no statistically significant difference between test implants and controls in marginal bone levels(P=.46) and RFA values. The mean BIC % in the Mg-implants was $54.5{pm}8.4%$. The mean BIC % in the turned implant was $45.3{pm}12.2%$. These differences between the Mg-implant and control implant were statistically significant(P=.005). CONCLUSIONS. The anodized, Mg-incorporated implant demonstrated significantly more bone-to-implant contact(BIC) in early loading conditions. CLINICAL IMPLICATIONS. The results of this study in beagle dogs suggest the possibility of achieving predictable stability of early loaded free-standing dental implants with Mg-incorporated surface.