- 헬멧형 뇌자도 장치의 센서 교정
- ㆍ 저자명
- 권혁찬,김기웅,유권규,김진목,이용호,Kwon. H.,Kim. K.,Yu. K.K.,Kim. J.M.,Lee. Y.H.
- ㆍ 간행물명
- Progress in superconductivity
- ㆍ 권/호정보
- 2010년|12권 1호|pp.57-61 (5 pages)
- ㆍ 발행정보
- 한국초전도학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
We have developed a whole-head MEG system for basic brain research and clinical application. The sensor system consists of a 152 SQUID gradiometer array oriented and located in a suitable way to cover a whole head of the human. The system measures magnetic fields generated by neuronal currents in the brain to get information on the brain activities. For this purpose, the field sensitivity determined by the position, orientation and geometry of the pickup coil as well as amplification factor of the electronic circuits should be known precisely. However, the position and orientation of the pickup coil might be changed from the designed specifications during cool down of the dewar and it is necessary to characterize the field sensitivity. In this study, we made calibration systems to determine the actual position and orientation of the 152 pickup coils and compared the localization results of the N100m source in the auditory cortex.