- 유방촬영의 화질관리 환경에 대한 고찰
- ㆍ 저자명
- 황인선,김영근,주형윤,Hwang. In-Sun,Kim. Young-Keun,Joo. Hyung-Yoon
- ㆍ 간행물명
- Korean journal of digital imaging in medicine : KJDIM
- ㆍ 권/호정보
- 2010년|12권 2호|pp.133-138 (6 pages)
- ㆍ 발행정보
- 대한디지털의료영상학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
The purpose of this study was to survey and test quality control of mammography system. The conclusion of this study is as follows ; First, The rate of pass for phantom image test shows that Film-Screen mammography system(F/S) and computed mammography system(CR) is 80%, Indirect digital mammography system(DR) is 100%. Second, The test of exposure dose shows that F/S is 921 mR. CR is 1,140 mR, DR is 474 mR. The grade of this testament is CR > F/S > DR. Third, The test of average glandular dose shows that F/S is 1,336 mGy, CR is 1,635 mGy, DR is 1,26 mGy. The grade of this testament is CR > F/S > DR. Fourth, The testament of resolution shows as follows F/S is 11~13 Lp/mm, CR is 4~5 Lp/mm, and DR is 5~7 Lp/mm(F/S > DR > CR) Fifth, The survey of projection, cassette, development and reading shows that user are indifference.