- 직.간접디지털 파노라마 방사선촬영시 흡수선량과 유효선량
- ㆍ 저자명
- 이근선,김진수,김재덕,Lee. Gun-Sun,Kim. Jin-Soo,Kim. Jae-Duk
- ㆍ 간행물명
- 대한구강악안면방사선학회지
- ㆍ 권/호정보
- 2010년|40권 1호|pp.9-14 (6 pages)
- ㆍ 발행정보
- 대한구강악안면방사선학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
Purpose : We evaluated the absorbed doses to the organs and calculated the effective doses when using the digital panoramic radiography. Materials and Methods : The absorbed dose averages in major organs of oral and maxillofacial region were measured using the Dental head phantom (CIRS Co., USA), $^nLi_2B_4O_7$ TLD chip and UD-716AGL dosimeter (Matsushita Electric Industrial Co., JPN) when performing indirect and direct digital panoramic radiography. Effective doses were calculated from correspond to ICRP 2007 recommendations for two panoramic radiography. Results : The absorbed dose average on indirect and direct digital panoramic radiography was highest in parotid glands as measured 1259.6 mGy and 680.7 mGy respectively. Absorbed dose average in another organs were high in order of esophagus, submandibular gland, tongue and thyroid gland on both types of digital panoramic radiography. The absorbed dose average was higher on indirect type than direct one (p<0.05). The effective dose was higher on indirect type than direct one as measured 13.28 mSv and 8.70 mSv respectively. Conclusion : The absorbed doses in salivary gland and oral mucosa were high. However, thyroid gland also demands the attention on radiography due to high tissue weighting factor in spite of the low absorbed dose.