Diagnostic reference levels in intraoral dental radiography in Korea
[STEP1]서지반출 형식 선택
파일형식
@
서지도구
SNS
기타
[STEP2]서지반출 정보 선택
  • 제목
  • URL
돌아가기
확인
취소
  • Diagnostic reference levels in intraoral dental radiography in Korea
  • Diagnostic reference levels in intraoral dental radiography in Korea
저자명
Kim. Eun-Kyung,Han. Won-Jeong,Choi. Jin-Woo,Jung. Yun-Hoa,Yoon. Suk-Ja,Lee. Jae-Seo
간행물명
Imaging science in dentistry
권/호정보
2012년|42권 4호|pp.237-242 (6 pages)
발행정보
대한구강악안면방사선학회
파일정보
정기간행물|ENG|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

기타언어초록

Purpose: The objectives of this study were to survey the radiographic exposure parameters, to measure the patient doses for intraoral dental radiography nationwide, and thus to establish the diagnostic reference levels (DRLs) in intraoral dental X-ray examination in Korea. Materials and Methods: One hundred two intraoral dental radiographic machines from all regions of South Korea were selected for this study. Radiographic exposure parameters, size of hospital, type of image receptor system, installation duration of machine, and type of dental X-ray machine were documented. Patient entrance doses (PED) and dose-area products (DAP) were measured three times at the end of the exit cone of the X-ray unit with a DAP meter (DIAMENTOR M4-KDK, PTW, Freiburg, Germany) for adult mandibular molar intraoral dental radiography, and corrections were made for room temperature and pressure. Measured PED and DAP were averaged and compared according to the size of hospital, type of image receptor system, installation duration, and type of dental X-ray machine. Results: The mean exposure parameters were 62.6 kVp, 7.9 mA, and 0.5 second for adult mandibular molar intraoral dental radiography. The mean patient dose was 2.11 mGy (PED) and 59.4 $mGycm^2$ (DAP) and the third quartile one 3.07 mGy (PED) and 87.4 $mGycm^2$ (DAP). Doses at university dental hospitals were lower than those at dental clinics (p<0.05). Doses of digital radiography (DR) type were lower than those of film-based type (p<0.05). Conclusion: We recommend 3.1 mGy (PED), 87.4 $mGycm^2$ (DAP) as the DRLs in adult mandibular molar intraoral dental radiography in Korea.