Morphologic Analysis of Water-Cooled Bipolar Radiofrequency lesions on Egg White in Vitro
[STEP1]서지반출 형식 선택
파일형식
@
서지도구
SNS
기타
[STEP2]서지반출 정보 선택
  • 제목
  • URL
돌아가기
확인
취소
  • Morphologic Analysis of Water-Cooled Bipolar Radiofrequency lesions on Egg White in Vitro
  • Morphologic Analysis of Water-Cooled Bipolar Radiofrequency lesions on Egg White in Vitro
저자명
Kang. Sang-Soo,Park. Jung-Chan,Yoon. Young-Jun,Shin. Keun-Man
간행물명
The Korean journal of pain
권/호정보
2012년|25권 3호|pp.151-154 (4 pages)
발행정보
대한통증학회
파일정보
정기간행물|ENG|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

기타언어초록

Background: The aim of this study was to document the optimal spacing of two cannulae to form continuous strip lesions and maximal surface area by using water-cooled bipolar radiofrequency technology. Methods: Two water-cooled needle probes (15 cm length, 18-gauge probe with 6 mm electrode tip) were placed in a parallel position 10, 20, 24, 26, and 28 mm apart and submerged in egg white. Temperatures of the probes were raised from $35^{circ}C$ to $90^{circ}C$ and the progress of lesion formation was photographed every 1 minute with the increase of the tip temperature. Approximately 30 photographs were taken. The resultant surface areas of the lesions were measured with the digital image program. Results: Continuous strip lesions were formed when the cannulae were spaced 24 mm or less apart; monopolar lesions around each cannula resulted if they were spaced more than 26 mm apart. Maximal surface areas through the formation of continuous strip lesion were 221 $mm^2$, 375 $mm^2$, and 476 $mm^2$ in 10, 20, and 24 mm, respectively. Summations of maximal surface area of each monopolar lesions were 394 $mm^2$ and 103 $mm^2$ in 26 and 28 mm, respectively. Conclusions: Water-cooled bipolar Radiofrequency technology creates continuous "strip" lesions proportional in size to the distance between the probes till the distance between cannulae is 24 mm or less. Spacing the cannulae 24 mm apart and treating about $80^{circ}C$ for 24 minutes maximizes the surface area of the lesion.