Comparison of effective dose for imaging of mandible between multi-detector CT and cone-beam CT
[STEP1]서지반출 형식 선택
파일형식
@
서지도구
SNS
기타
[STEP2]서지반출 정보 선택
  • 제목
  • URL
돌아가기
확인
취소
  • Comparison of effective dose for imaging of mandible between multi-detector CT and cone-beam CT
  • Comparison of effective dose for imaging of mandible between multi-detector CT and cone-beam CT
저자명
Jeong. Dae-Kyo,Lee. Sang-Chul,Huh. Kyung-Hoe,Yi. Won-Jin,Heo. Min-Suk,Lee. Sam-Sun,Choi. Soon-Chul
간행물명
Imaging science in dentistry
권/호정보
2012년|42권 2호|pp.65-70 (6 pages)
발행정보
대한구강악안면방사선학회
파일정보
정기간행물|ENG|
PDF텍스트
주제분야
기타
이 논문은 한국과학기술정보연구원과 논문 연계를 통해 무료로 제공되는 원문입니다.
서지반출

기타언어초록

Purpose : The aim of this study was to compare the effective dose for imaging of mandible between multi-detector computed tomography (MDCT) and cone-beam computed tomography (CBCT). An MDCT with low dose technique was also compared with them. Materials and Methods : Thermoluminescent dosimeter (TLD) chips were placed at 25 organ sites of an anthropomorphic phantom. The mandible of the phantom was exposed using 2 different types of MDCT units (Somatom Sensation 10 for standard-dose MDCT, Somatom Emotion 6 for low-dose MDCT) and 3 different CBCT units (AZ3000CT, Implagraphy, and Kavo 3D eXaM). The radiation absorbed dose was measured and the effective dose was calculated according to the ICRP 2007 report. Results : The effective dose was the highest for Somatom Sensation 10 (425.84 ${mu}Sv$), followed by AZ3000CT (332.4 ${mu}Sv$), Somatom Emotion 6 (199.38 ${mu}Sv$), and 3D eXaM (111.6 ${mu}Sv$); it was the lowest for Implagraphy (83.09 ${mu}Sv$). The CBCT showed significant variation in dose level with different device. Conclusion : The effective doses of MDCTs were not significantly different from those of CBCTs for imaging of mandible. The effective dose of MDCT could be markedly decreased by using the low-dose technique.