- 개인용 혈당측정기의 신뢰성 보증시험 개발
- ㆍ 저자명
- 김기영,박호준,박찬수,함중걸,장중순,Kim. Gi-Young,Park. Ho-Joon,Park. Chan-Soo,Ham. Jung-Keol,Jang. Joong Soon
- ㆍ 간행물명
- 신뢰성응용연구
- ㆍ 권/호정보
- 2013년|13권 4호|pp.299-307 (9 pages)
- ㆍ 발행정보
- 한국신뢰성학회
- ㆍ 파일정보
- 정기간행물| PDF텍스트
- ㆍ 주제분야
- 기타
The value of blood sugar is measured by a personal blood glucose meter which consists of two parts: a glucose strip and a blood glucose meter. A personal blood glucose meter makes use of electric media which are composed of glucose oxidase and electron. This study is to propose a method of reliability demonstration test derived from field data analysis and FMEA(Failure Mode and Effect Analysis). Detailed Conditions for reliability demonstration test are selected from the comparison of various failure mechanisms. The most dominant failure mechanism is wear-out which is caused by strip insertion/extraction. The testing device that can reproduce the failure mechanism of strip insertion/extraction is made to conduct reliability demonstration test. Using the testing device, it is confirmed that target lifetime of selected devices is more than 2 years.